Output Signal |
Standard 2-wire: 4-20 mA / Vs = 12-36 Vdc (CSA) |
Accuracy |
<±0.35% FSO BFSL
Optional nominal pressure > 5 psi: <±0.25% FSO BFSL
|
Permissible Load |
Rmax = [(Vs – Vs min) / 0.02] Ohm |
Influence Effects |
Supply: 0.05% FSO / 10 V
Load: 0.05% FSO / kOhm
|
Long-term Stability |
<±0.1% FSO/yr |
Response Time |
<10 ms |
Permissible Temperatures |
Medium: 14°F to 158°F (-10°C to 70°C)
Storage: -13°F to 158°F (-25°C to 70°C)
|
Reverse Polarity Protection |
No damage. No function. |
Electromagnetic Compatibility |
Emission and immunity according to EN 61326 |
Electrical Connection |
Cable with sheath material*: PVC grey, PUR black, FEP black |
Housing |
316Ti SS |
Seals |
FKM / EPDM |
Diaphragm |
316L SS |
Weight |
100 g (without cable) |
Current Consumption |
25 mA max |
Connecting Cables |
Cable capacitance: Signal line/shield and signal line/signal line: 160 pF/m
Cable inductance: Signal line/shield and signal line/signal line: 1 µH/m
|
Approval |
CSA Certified for electrical equipment |
Enclosure Rating |
NEMA 6P / IP68 |
*Cable with integrated air tube for atmosphere pressure reference |